Experimental Evaluation of SAR around an Implanted Cardiac Pacemaker Caused by Mobile Radio Terminal

Yuta ENDO
Kazuyuki SAITO
Soichi WATANABE
Masaharu TAKAHASHI
Koichi ITO

Publication
IEICE TRANSACTIONS on Communications   Vol.E95-B    No.6    pp.2129-2132
Publication Date: 2012/06/01
Online ISSN: 1745-1345
DOI: 10.1587/transcom.E95.B.2129
Print ISSN: 0916-8516
Type of Manuscript: LETTER
Category: Electromagnetic Compatibility(EMC)
Keyword: 
implanted cardiac pacemaker,  SAR,  phantom,  FDTD,  PIFA,  

Full Text: PDF(679.5KB)>>
Buy this Article



Summary: 
Although the effect of electromagnetic interference on an implanted cardiac pacemaker due to a nearby mobile phone has been investigated, there have been few studies on the enhancement of the specific absorption rate (SAR) around an implanted cardiac pacemaker due to a nearby mobile phone. In this study, the SAR distribution around a pacemaker model embedded in a parallelepiped torso phantom when a mobile phone was nearby was numerically calculated and experimentally measured. The results of both investigations showed a characteristic SAR distribution. The system presented can be used to estimate the effects of electromagnetic interference on implanted electric circuits and thus could lead to the development of guidelines for the safe use of mobile radio terminals near people with medical implants.